Ⅱ. Technical Specification
- Range and Accuracy Error
|
Measurement function
|
Measuring range
|
Accuracy
|
Resolution
|
|
ground resistance
(R)
|
0.00Ω~30.00Ω
|
±2%rdg±5dgt (Note 1)
|
0.01Ω
|
|
30.0Ω~300.0Ω
|
±2%rdg±3dgt
|
0.1Ω
|
|
300Ω~3000Ω
|
±2%rdg±3dgt
|
1Ω
|
|
3.00kΩ~30.00kΩ
|
±2%rdg±3dgt
|
10Ω
|
|
Soil resistivity
(ρ)
|
0.00Ωm~99.99Ωm
|
ρ=2πaR (Note 2)
|
0.01Ωm
|
|
100.0Ωm~999.9Ωm
|
0.1Ωm
|
|
1000Ωm~9999Ωm
|
1Ωm
|
|
10.00kΩm~99.99kΩm
|
10Ωm
|
|
100.0kΩm~999.9kΩm
|
100Ωm
|
|
1000kΩm~9999kΩm
|
1kΩm
|
|
ground voltage
|
AC 0.00~100.0V
|
±2%rdg±3dgt
|
0.01V
|
Note: 1. Reference condition: Accuracy when Rh Rs<100Ω.
Working conditions: Rh max=3kΩ+100R<50kΩ; Rs max=3kΩ+100R<50kΩ
- It depends on the measurement accuracy of R, π=3.14, a: 1 m~100m;
- General Specifications
|
Features
|
Two-three-four-wire measurement of grounding resistance and soil resistivity;
Ground voltage measurement
|
|
Ambient temperature and humidity
|
23℃±5℃,below 75%rh
|
|
Power supply
|
DC 6V 4.5Ah lead-acid battery for 100 hours of continuous standby
|
|
Interference voltage
|
<20V(should be avoided)
|
|
Disturbance current
|
<2A(should be avoided)
|
|
Electrode spacing when measuring R
|
a>5d
|
|
Electrode spacing when measuring ρ
|
a>20h
|
|
Auxiliary grounding resistance value
|
Reference condition <100Ω, working condition <5kΩ
|
|
Range
|
Grounding resistance: 0.00Ω~30.00kΩ
|
|
Soil resistivity: 0.00Ωm~9999kΩm
|
|
Ground voltage: 0.00V~100.0V
|
|
Measurement method
|
Precision 4-wire, 3-wire method measurement, simple 2-wire measurement ground resistance
|
|
Measurement methods
|
Grounding resistance: rated current pole change method
Soil Resistivity: Quadrupole Method
Ground voltage: average value rectification (between S-ES interface)
|
|
Test frequency
|
128Hz
|
|
Short circuit test current
|
> 20mA (sine wave)
|
|
Open circuit test voltage
|
AC 28V max
|
|
Electrode Spacing Range
|
Can be set from 1m to 100m
|
|
Shift
|
Grounding resistance: 0.00Ω~30.00kΩ automatic gear shifting
|
|
Soil resistivity: 0.00Ωm~9999kΩm automatic shifting
|
|
Backlight
|
Controllable gray and white backlight, suitable for use in dark places
|
|
Display mode
|
4-digit large LCD display, gray and white backlight
|
|
Measurement instructions
|
LED blinks during measurement
|
|
LCD size
|
111mm×68mm
|
|
LCD display field
|
108mm×65mm
|
|
Meter size
|
Length, width and height: 277.2mm×227.5mm×153mm
|
|
Standard test lead
|
4 strips: 15m for red, 15m for black, 10m for yellow, 10m for green
|
|
Simple test lead
|
2: Yellow 1.6m, 1 green 1.6m each
|
|
Auxiliary ground rod
|
4 pieces: φ10mm×200mm
|
|
Measure time
|
Voltage to ground: about 3 times/second
|
|
Grounding resistance, soil resistivity: about 7 seconds/time
|
|
Ground voltage
|
AC 100V or less measurement (the ground voltage measurement function cannot be used to measure commercial power)
|
|
USB interface
|
With USB interface, software monitoring, storage data can be uploaded to the computer, save and print
|
|
Communication line
|
1 USB communication cable, 1.5m long
|
|
Data storage
|
500 groups, "MEM" storage indication, displaying "FULL" symbol means the storage is full
|
|
Data access
|
"MR" symbol indication when viewing data
|
|
Overflow display
|
"OL" symbol indication when overrange overflow
|
|
Interference test
|
Automatically identify interference signals, and the "NOISE" symbol indicates when the interference voltage is higher than 5V
|
|
Auxiliary ground test
|
With auxiliary grounding resistance value test function, 0.00KΩ~30kΩ (Rh max=3kΩ+100R<50kΩ; Rs max=3kΩ+100R<50kΩ)
|
|
Alarm function
|
When the measured value exceeds the alarm setting value, an alarm prompt will be issued
|
|
Battery voltage
|
Real-time display of battery power, reminding to charge in time when the battery voltage is low
|
|
Automatic shut-down
|
The meter shuts down after about 15 minutes of inactivity
|
|
Power consumption
|
Standby: 40mA Max (backlight off)
|
|
Turn on the backlight: 43mA Max
|
|
Measurement: 120mA Max (backlight off)
|
|
Quality
|
Meter: 2406 (including battery)
|
|
Test line: 1300g
|
|
Auxiliary ground rod: 850g (4 pieces)
|
|
Working temperature and humidity
|
-10℃~40℃; below 80%rh
|
|
Storage temperature and humidity
|
-20℃~60℃; below 70%rh
|
|
Overload protection
|
Measuring ground resistance: AC 280V/3 seconds between each port of H-E and S-ES
|
|
Insulation resistance
|
Above 20MΩ (500V between circuit and case)
|
|
Pressure resistance
|
AC 3700V/rms (between circuit and case)
|
|
Electromagnetic properties
|
IEC61326(EMC)
|
|
Suitable for safety regulations
|
IEC61010-1 (CAT III 300V, CAT IV 150V, pollution degree 2);
IEC61010-031;
IEC61557-1 (grounding resistance);
IEC61557-5 (soil resistivity);
JJG 366-2004.
|
Ⅲ. Packing List
|
Meter
|
1 set
|
|
Instrument bag
|
1
|
|
Auxiliary ground rod
|
4 sticks
|
|
Standard test lead
|
4 strips (15 meters for red; 10 meters for yellow; 10 meters for green; 1 for each of 15 meters for black)
|
|
Simple test lead
|
2 strips (yellow 1.6 meters; green 1.6 meters)
|
|
6V lead-acid battery (built-in)
|
1
|
|
Charger
|
1
|
|
Monitoring software CD
|
1 serving
|
|
USB communication cable
|
1
|
|
Manual, Warranty
|
1 set
|
Condition
PS:What you see is what you get. After receiving the goods, please take an unpacking video for record. If there is any issue with the product, you can return or exchange it within 15 days.
*** Payment ***
Orders must be paid for within 5 Business days.
Once payment is received & verified, we will send out the shipment within 5 business days, excepting holidays.
*** Order Processing & Handling Time ***
Your order will be shipped within 5 business days (excluding Friday, Saturday, and Jewish Holidays) following confirmation of payment.
This handling time is needed for special international secure packaging and paperwork.
You will be informed by Email of any unexpected shipping delays, should the situation warrant.
*** Delivery Time ***
Your order will be shipped from China Guangzhou (Registered mail with tracking).
Full tracking information will be provided to you.
Once your order is shipped, it normally takes : ( Please read the purchase description)
Regular Air Mail (Economy - Post Office ) : 11-30 business days
EMS (Standard) : 8-15 days for door-to-door delivery.
UPS /DHL (Buyer Request ): 6-10 days
N.B. ("note well") These shipping times are estimations only. Receipt of shipment may take longer, depending on the local situation in your country. Delivery time varies according to destination.
Free shipping within the United States. For remote areas and countries in South America, please consult our customer service to find out whether the item can be shipped. There may also be a possibility of additional shipping fees.